Take another look at the Swissplus - Ultimate convenience and simplicity.

A single stage implant system that comes with everything you need to streamline your implant cases. Swissplus Straight and Tapered implants help minimize chair time and financial investment by utilizing a unique, all-in-one package that includes the implant, healing collar and a patented Fixture Mount/Transfer that functions as an impression post and a final abutment. With no extra parts to buy, Swissplus is a versatile and complete single-stage solution. Tapered Swissplus, with its double-lead threads enhances primary stability, making it ideal for use with immediate implant techniques - further simplifying your implant cases.
Technical Information - Implant properties and Dimensions
Implant Diameter:
- Straight Swissplus Implants are available in two body diameters, 4.1mmD and 4.8mmD.
- Tapered Swissplus Implants are available in two body diameters, 3.7mmD and 4.8mmD.


Implant platforms:
- 4.8mmD platform - (Fig. 1a). A 45° external beveled shoulder tapers up from the height of contour of the implant to the coronal area, which has a 3.5mmD opening. The height of this bevel is approximately 0.6mm above the height of contour. From the edge of the opening, an 8° tapered internal beveled wall leads into a 3.0mmD flat-to-flat octagon, which is 1.5mm deep. Below the octagon is a continuation of the 8° bevel, which leads into the threaded area where the fixation screw is received.
- 3.8mmD platform - (Fig. 2a & b). A 40° external beveled shoulder tapers up from the height of contour of the implant to the coronal area, which has a 2.85mmD opening. The height of this bevel is approximately 0.5mmL above the height of contour. From the edge of the opening, a 15° tapered internal beveled wall leads into a 2.5mmD flat-to-flat hexagon, which is 1.85mm deep. Below the hexagon is the threaded area where the fixation screw is received.

Implant material:
Cold worked Grade 4 commercially pure titanium has an Ultimate Tensile Strength of 130 ksi (896 MPa) withstands 23 in-lbs (260 Ncm) of torque and withstands between 372-753 lbs (1655N-3350N) of compression at 30°.

Implant thread design:
Tapered SwissPlus Implant with double lead thread
Fig. 2a) Side view of both threads, colored red or blue.
Fig. 2b) Side view of implant with one thread colored to indicate the presence of two threads traversing the length of the threaded portion of the implant.
Fig. 2c) Apical end of the Tapered SwissPlus indicating the start of the two threads 180° from each other.

Straight SwissPlus Implant with single lead thread
Fig. 1a) Side view of thread.
Fig. 1b) Apical end of the Straight SwissPlus indicating the start of the single thread.